Hyperbaric outlet boxes are devices used to connect medical equipment, such as hyperbaric chambers, to a building’s medical gas supply system. They are typically used in hospitals and other healthcare facilities to provide a safe and reliable supply of medical gases, such as oxygen and nitrous oxide, for hyperbaric oxygen therapy (HBOT) and other treatments.
Hyperbaric outlet boxes typically consist of a wall-mounted plate with one or more outlets, as well as a valve that controls the flow of gas. They are designed to fit a specific connector style, such as the DISS style connector, which is commonly used in hospitals. The outlet boxes are connected to a hyperbaric manifold or another central distribution system, which is used to distribute medical gases throughout the facility.
It is important to use hyperbaric outlet boxes according to the manufacturer’s instructions and to follow all safety precautions to prevent gas leaks or other hazards. It is also important to regularly maintain and inspect hyperbaric outlet boxes to ensure that they are functioning properly and to address any issues that may arise.
We have designed the perfect solution for connecting medical gas and vent lines to hyperbaric chamber installations the Hyperbaric Valve/Outlets/Vent Box. The HCB Box is designed in a recessed box with single point plumbing connections for fast easy installation, a high flow (5,000 scfh / 2,360 lpm) oxygen outlet and with a professional cosmetic look to make your hyperbaric installation state-of-the art.
Hyperbaric Outlet Boxes Features:
- Valves are forged brass (not cast)
- All valves are dual gauge port full portball valves
- 3 piece in-line repairable valves
- Adjustable packing with blow-out proof stem design
- Cleaned for Oxygen service
- Plastic insulators to protect against galvanic corrosion
- Provided with 1 ½” diameter gauges
- Outlets conform to CGA standards (except high flow Oxygen outlet)
- Conforms to NFPA 99 & CGA standards (where applicable)
- Outlets accept DISS gas specific nut & nipple sets (except “OH” high flow oxygen model – see below
- Ease of maintenance without removing the faceplate using a cartridge insertion tool on outlets.
- All outlets are 100% tested for flow and leaks
- Manufactured & Tested in the U.S.A. with Domestic & Imported Components